Archives of Acoustics, 15, 3-4, pp. 355-365, 1990

Digital velocity profile estimator of blood flow in the heart and large blood vessels

P. Karłowicz
Institute of Fundamental Technological Research, Polish Academy of Sciences, 00-049 Warszawa, ul. Świętokrzyska 21
Poland

J. Liwski
Institute of Fundamental Technological Research, Polish Academy of Sciences, 00-049 Warszawa, ul. Świętokrzyska 21
Poland

M. Piechocki
Institute of Fundamental Technological Research, Polish Academy of Sciences, 00-049 Warszawa, ul. Świętokrzyska 21
Poland

W. Secomski
Institute of Fundamental Technological Research, Polish Academy of Sciences, 00-049 Warszawa, ul. Świętokrzyska 21
Poland

A. Nowicki
Institute of Fundamental Technological Research, Polish Academy of Sciences, 00-049 Warszawa, ul. Świętokrzyska 21
Poland

The paper discusses principles of a digital multigate system for Doppler frequency measurements. The conception of a system with serial data processing, based on the Doppler frequency measurement with the "zero crossing" method, is described. Problems with the recognition of the direction of flow in the digital system are presented, as well as two so]utiolls, tested in practice, whjch limit occurring measuring errors. An example of an actua] flow measurement made with the application of a model of the presented device is described, and possible errors resulting from the "zero crossing" technique are given. In order to improve properties of the system the use of a digitallow-pass filter is proposed. The conception of such a filter is described.
Full Text: PDF