Archives of Acoustics, 15, 3-4, pp. 375-397, 1990

Design of ultrasonic probes for medical diag¬nostics

G. Łypacewicz
Ultrasonic Department, Institute of Fundamental Technological Research, Polish Academy of Sciences, 00-049 Warszawa, Świętokrzyska 21
Poland

E. Duriasz
Ultrasonic Department, Institute of Fundamental Technological Research, Polish Academy of Sciences, 00-049 Warszawa, Świętokrzyska 21
Poland

The Mason's equivalent circuit of the piezoelectric transducer is modified in order to investigate separately the influence of parameters of piezoelectric material as well as electrical and acoustical elements applied in the ultrasonic probe. All values transfered to the mechanical side of the electromechanical transformer let us introduce the relative values both electrical and acoustical parameters. The mechanical and dielectric losses of the transducer are taken into consideration.
The influence of various parameters of the probe on input admittance and reflected from an ideal reflector pulses is shown.